SPECIFICATIONS
Combin ation Blan ket/Flui d W armer Operat or s Manua l MN -399 40 Rev 3 8/20
17
EC1350BL Built-In Cutout Specifications and Inlet
Locations
Continued on next page
CAUTION: Fire hazard.
Supply wires must be rated for a minimum wire temperature of
80°C.
NOTICE
Built-in units must be hardwired to the facility mains with a 20A breaker switch.
Breaker* Wire Size** Wire Temperature
20A 12 AWG (1mm
2
) 80°C (minimum)
*Electrical connections must meet applicable codes.
**IEC 60335 minimum.
Cutout Specifications Inlet Options
Height 74.5” (1892.25mm) 1 Back of warmer
Width 29.25” (743mm) 2 Top of warmer
Depth 17” (432mm) 3 Left side of warmer
Top of unit clearance 12” (304.75mm)