OPERATION
Combinat ion Blan ke t/F luid W a rmer ▪ Operato r’s Man ual ▪ M N-399 40 ▪ Rev 3 ▪ 8/ 20
32
Continued from previous page
2. Set the temperature using the arrow buttons .
3. Open the door and load the fluid or blankets into the warmer.
Close the door.
During the warming
process
4. Press the temperature recall button to view the measured cavity
temperature.
The measured cavity temperature will display for five seconds. Then, the
set-point temperature will display.
Result The fluids or blankets are now warming.
NOTE: The irrigation temperature set-point range is
90°F – 150°F (32°C – 66°C).
The injection temperature set-point range is 90°F – 104°F
(32°C – 40°C).
The blanket temperature set-point range is 90°F – 160°F
(32°C – 71°C).