TROUBLESHOOTING
Combin ation Blan ket/Flui d W armer Operat or s Manua l MN -399 40 Rev 3 8/20
49
E-60 Real Time Clock
Checksum Error
(Blanket warmers only)
Real time clock rechargeable battery backup has discharged.
Plug the warmer into the outlet for 30 minutes.
*E-61 Real Time Clock
(Blanket warmers only)
Real time clock not responding. Contact service if error persists.
E-62 Real Time Clock
(Blanket warmers only)
Timer overlay is present, but no real time clock is detected. Contact service.
*E-70 Pad Count Error More heater pads detected than set for. Hold the standby button for 12
seconds until display shows "PAd#” (# = number of pads selected [3-7]).
Press up or down arrow to adjust to correct number of pads. [Blanket
warmers: 1.5ft
3
, 2.5ft
3
= 3 pads, 3.5ft
3
, 4.0ft
3
, 7.5ft
3
= 4 pads.
Fluid warmers: 2.5ft
3
& 4.0ft
3
= 3 pads].
*E-71 Personality Error Contact service.
E-80 EEPROM Error EEPROM not responding. Contact service if error persists.
*E-81 Calibration not
locked
Contact service.
*E-83 EEPROM Error Contact service for help resetting the control.
E-87 EEPROM Error Stored offsets corrupted. Offsets reset to 0.
Control may need to be re-calibrated.
Possible bad EEPROM.
Contact service if error persists.
E-90 Button stuck A button has been held down for >60 seconds. Adjust control. Error will
reset when the problem has been resolved.
E-95 Factory Test pin
short detected.
Ensure that debris is not causing a short between the test pins. If the pins
are clean, replace the control.
*E-98 Temperature
Delta Error
Temperature of the cavity sensors 1 and 2 differ by more than 3.3ºC (6ºF).
Remove product and allow the warmer to cool down.
Verify that the product sensor is clean and operating correctly.
Set the power switch to the OFF position to clear the error code.
If error persists, the cavity sensor should be replaced by a qualified
service technician. Contact service.
E-99 Hardware Over
Temp
Inspect the connections and condition of high limit bimetal thermostat.
If applicable, make sure that the compartment fan motor is operating.
Contact service if error persists.