Can be adjusted to between
SpO2:
Upper limit: 51 % - 100%
Lower limit: 50 % - 99%
Pulse:
Upper limit: 31 - 250 BPM
Lower limit: 30 – 249 BPM
Sound frequency variable (depending on measured SpO2 value)
Max. 96 h with a storage interval of 10 s
(Storage interval can be changed from 5 s to 60 s)
Measurement values, date, time, alarm limits, events
Unit and data in connection with MySign
®
software
(e.g. name, ward, patient ID)
This unit complies with the requirements of MDD 93/42/EEC concerning
medical devices and the corresponding standards.
It also complies with: DIN EN 1789 Medical vehicles
and their equipment - Ambulances
* with SoftTip
®
R-3211-12 MySign
®
Subject to technical changes!
* General accuracy statement - refer to the sensor specific clinical accuracy described below for detailed information.
066-07-1002197_GA_MySignS_FDA / 06.14
45