13 Technical specifications
All of the specifications apply to the following standard conditions: Ambient temperature of 1013 hPa, 25°C dry
ambient air.
Saturation (SpO2) 1 to 100%
Pulse rate (PR) 0 to 300 BPM
Pulsation index (PI) below 0.1 to 20
Saturation +/- 2% (70 to 100%, without disrupting movement)
Pulse rate +/- 3 BPM (30 to 250 BPM, without disrupting movement)
Time interval for updating the data on the display: 300 ms
Plethysmogram: Amplitude = 75% of the display area
30 – 95% rel. humidity (non-condensing), for operation outside this range
refer to the safety information Risk of Electrostatic Discharge
Operating time per charge
> 18 hours (standard settings)
> 24 hours with auto-off activated
Function with the lowest display brightness setting
Mini-USB type B, protection class II
Input: AC 110V - 230V / 50 – 60 Hz / 125 mA
Output: DC 5V / ≥ 1 A / < 15 W
IEC 60601-1 / IEC 60950-1
Full functionality guaranteed
160 x 72 x 39 mm (L x W x H)
Approx. 320 g (with sensor*)
USB 2.0 (socket Mini-USB type B)
Monitoring the alarm limits and unit functions (visual and audio)