6.6 Programming the recorder
After you have selected the patient, a window will appear for configuring
and programming the PhysioQuant Recorder.
Number of measurement intervals
Up to 4 different measurement intervals can be selected. The standard
setting is at 2 intervals (day and night phase).
Power supply
The type (normal battery or rechargeable battery) that is used for long-
term measurements is set.
From ... To ...
The beginning and end of selected intervals can be set in minutes.
Measurement interval
Exact: measurements are taken at the exact intervals in minutes that
have been set.
Approx.: measurements randomly vary at approx. +/- 2 minutes around
the times set.
Minutes
The interval between subsequent measurements can be programmed to
range between 2 and 90 minutes.