6.10 Important measurement information
For the first measurement, the device will inflate the cuff to approx. 160
mmHg (starting pressure). For subsequent measurements, the inflation
pressure will be 25 mmHg above the last measured systolic value
(minimum 120mmHg).
If the measurement value is in excess of the inflation pressure, then
additional inflation of 50 mmHg will take place.
A manual measurement can be taken at any time between automatic
measurements. This will be indicated by '+' in the measurement value
table.
After an incorrect measurement, the measurement will be repeated after
2 minutes. An error message is only displayed in the error table with its
corresponding error code after 3 incorrect measurements.
After the error messages 'E04' (Empty battery/rechargeable battery),
'E07' (Inflation time finished) and 'E10' (200 measurements carried out),
the measurement will not be repeated. After error message 'E07', the
subsequent measurement will again take place according to the selected
measurement cycle.
After error messages 'E04' and 'E10' have been displayed, the device will
switch to a power-saving mode, in order to avoid complete discharge of
the rechargeable batteries. This mode can only be cancelled by
switching the device on and off.