7 Maintenance and servicing
80116-941_V24392
2023-03
Test procedure
1. Test step
(filled bottle)
1. Set the test specimen to:
APC probe, monopolar – forcedAPC – Effect 0.1
2. In the bottle fill level window, compare the fill level display (%) with
the display on the pressure regulator manometer. Both should essen-
tially match.
IMPORTANT! The maximum bottle pressure can be set to between 100 and 240 bar
(see <Further settings -> Argon plasma module APC>). For correct dis
-
play of the bottle fill level on the VIO 3, the bottle pressure setting of
the unit must match the bottle pressure of the test bottle.