7 Maintenance and servicing
80116-941_V24392
2023-03
4. Call up <Protected settings>.
5. Enter the password Level2 and confirm with <Entry>. The unit shows
all password-protected settings.
6. The following settings are relevant:
Use of AUTO START: Must be permitted for the test.
Delay AUTO START thermoSEAL: Note the value.
7. Exit <Further settings>.
8. Set the test specimen to:
BiClamp, bipolar– thermoSEAL – Effect 1 –AUTO START
9. Hold down the S1 button on the VIO test box until the test specimen
is activated. Activation must begin after the test delay that has been
set.
10. Press the S2 button. The test specimen must terminate the activation.