Model-No.: JK 155 / 46-3 AC
Power consumption rating: 13,700 W
Rated frequency: 60 Hz
Rated voltage: 230 V 3 phase 230 V 2 phase
Rated overcurrent protection device: 50 amp 3-pole 80 amp 2-pole
Power supply: 3 x THHN AWG 8 or 6 x THHN AWG 8
Classification: Type BF applied part
Protection ingress classification: IP20
Protection against shock: Class I
Max. allowed network impedance: 0.095
Noise level 1m away from device:
(without exhaust air):
67.2 dB(A)
Noise level inside the device: < 70 dB(A)
Load on the acrylic panel: max. 330 lb / 150 kg
Weight: approx. 1,327 lb / 602 kg
Max. humidity: < 70%
Ambient temperature: 59-89 °F (15-32 °C)
Max. operating altitude (above sea level): < 6500 ft (2000 m)
NOTE:
Adaptation of the unit is required for operating altitudes
greater than 6,500 ft / 2,000 m above sea level in order to
guarantee trouble-free operation.
Essential performance
NOTE:
The essential performance of the device is the UV exposure
dose within specified limits, whereas the upper limit must not
be exceeded. No-radiation or zero radiation does not result or
contribute to an injury of the user.
Lower limit exposure dose 0,75 MED
*)1
Upper limit exposure dose 4 MED
*)
1
*) in compliance with 21 CFR 1040.20 and FDA’s timer policy