EasyManua.ls Logo

ergoline AFFINITY 700 - Page 87

ergoline AFFINITY 700
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A
ppendi
x
10/2020 1020217-00A Operating instructions – 87/96
16.1.3 Table 204: Guidelines and manufacturer's
declaration – Electromagnetic interference
The Tanning system has been designed for use in the electromagnetic envi-
ronment specified below. The device operator must ensure that it is used in
such an environment.
Interference
immunity tests
IEC 60601 -
Test level
Defined
level
Electromagnetic environment – Guidelines
Conducted HF noise
IEC 61000-4-6
3 V rms 150 kHz to 80 MHz
outside the ISM band, 6 V
RMS in the ISM and ama-
teur radio bands
Modulation:
80% AM at 1kHz
3 V rms 150 kHz to 80 MHz
outside the ISM band, 6 V
RMS in the ISM and ama-
teur radio bands
Modulation:
80% AM at 1kHz
Portable and mobile radios are not to be used at shorter distanc-
es to the Model-No.: JK 155 / 46-3 and its cables other than the
recommended safe distance, which has been calculated accord-
ing to the transmitting frequency are made.
Recommended safe distance:
80 MHz to 800 MHz
800 MHz to 2,7 GHz
P is the transmitter's maximum output power (nominal rating) in
Watts (W) according to the details provided by the manufacturer
of the transmitter and the recommended safe distance is given in
metres (m).
The field strength of the stationary radio transmitter is for all
frequencies
b
in accordance with tests that were carried out on-
site
a
less than the determined level.
Interference is possible in the immediate surrounding of equip-
ment that carry the following symbols.
Radiated HF noise
IEC 61000-4-3
10 V/m 80 MHz to 2,7GHz
Modulation:
80% AM at 1kHz
27 V/m 385 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 450 MHz
Modulation:
FM + 5 Hz deviation:
1 kHz sine
9 V/m 710, 745, 780 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 810, 870, 930 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 1720, 1845 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 1970, 2450 MHz
Modulation:
Pulse Modulation: 217 Hz
9 V/m 5240, 5500,
5785 MHz
Modulation:
Pulse Modulation: 217 Hz
10 V/m 80 MHz to 2,7GHz
Modulation:
80% AM at 1kHz
27 V/m 385 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 450 MHz
Modulation:
FM + 5 Hz deviation:
1 kHz sine
9 V/m 710, 745, 780 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 810, 870, 930 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 1720, 1845 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 1970, 2450 MHz
Modulation:
Pulse Modulation: 217 Hz
9 V/m 5240, 5500,
5785 MHz
Modulation:
Pulse Modulation: 217 Hz
Comment 1: The higher value applies to 80 MHz und 800 MHz.
Comment 2: These guidelines might not apply to all situations. The propagation of electro-
magnetic waves is affected by absorption and reflexes caused by buildings, objects and peo-
ple.
a The field strength of a stationary transmitter e.g. standard cell phone stations (mo-
bile / cordless phones) and public mobile radio services, amateur stations, MW and UHF radio
and TV transmitters) cannot be precisely determined for theoretical reasons. It is recommend-
ed that an onsite test should be carried out in order to determine the electromagnetic environ-
ment caused by stationary HF transmitters. If the field strength of the massage device deter-
mined at the installation site exceeds the defined HF level listed above then the device must
be monitored with regard to its normal operation. If any abnormal characteristics are recorded
then it will be necessary to introduce additional measures, such as reorientation or converting
the device.
b The field strength must be less than 3 V/m over the frequency ranges from 150 kHz
to 80 MHz.