33
Class II Biological Safety Cabinets
3. Overlap the tape at the joints. 4. These 2 cables (for BSC and for vaporizer) should
be combined so that only 1 cable should come
out from the decontamination bag, to minimize
chance of formalin leakage.
5. To seal the cable, first make an Ω-shape
aluminum tape around the cable, and make a 5
cm flat portion going to both directions. Then,
paste the aluminum tape to the floor.
6. Put the bottom part of the decontamination bag
on top of the aluminum tape, and then seal it
with another layer of aluminum tape. Use
multiple layer of aluminum tape (at least 2 layers)
on both directions to ensure leak tightness.
7. The BSC is ready for formalin
decontamination.