EasyManua.ls Logo

Esco Multi-zone ART Workstation - O2 Sub-Menu (Only MAW with a Built-In Gas Mixer)

Esco Multi-zone ART Workstation
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAW
®
User Manual Rev. 3.0
32
3. Adjust the calibration to the desired level by pressing (⇧) or (⇩). In this case, we want to
adjust the value to 5.4%. Press (⇧) 4 times quickly. Sensor reading has delay, so response
time may be slower.
4. Release the SP key. The value is now stored and the CO
2
sensor calibration has been mod-
ified.
13.3.5 O
2
sub-menu (only MAW
®
with a built-in gas mixer)
Press (⇨) on O
2
to enter the O
2
sub-menu.
First item in the O
2
sub-menu is O
2
sensor calibration.
Calibrate holding down the SP key; use (⇧) and (⇩) to adjust.
Move to the next sub-menu item with (⇩) or one step up with (⇧).
Toggle O
2
regulation ON/OFF by holding the SP key and pressing (⇧) or (⇩).
The default status for the O
2
control is OFF.
Press (⇩) to move to the next item in the O
2
sub-menu.
N
2
flow rate is shown. This is the amount of N
2
gas put into the system while regulating. The
volume is shown in l/hour. It cannot be adjusted. Normally it will fluctuate along with the
O
2
regulation.
Press (⇩) to move to the next item in the O
2
sub-menu:
N
2
internal pressure rate is shown. The value is in bar. It must be between >0.2 and ≤0.8 bar
(2.90 11.60 PSI) at all times. It cannot be adjusted on the incubator it must be adjusted
on the external gas regulator.
Example how to calibrate the O
2
:
Using a suitable and calibrated device on one of the gas sample ports, the real O
2
concentra-
tion has been recorded to be 5.3%. Both ports should be used during calibration.