25 User troubleshooting
Table 25.1 Heating system
No heating, display is off
The unit is switched off at the back or
not connected to the power source
Switch on the device or connect to
the power source
Temperature deviated more than
0.5
o
C from the set temperature
Temperature setpoint is incorrect
Check the desired temperature
setpoint
Calibrate each zone according to
the User manual, using a high-pre-
cision thermometer
‘CO2 P’ is shown on the dis-
play.
No/wrong CO
2
gas pressure in the
system
Check CO
2
gas supply; make sure
that the pressure is kept stable at
0.6 bar (8.70 PSI)
The system is not powered
The system is on standby or
switched off
The data cable between incubator
and PC is not properly attached
Check the connection. Use only
the cable supplied with the unit
Missing segment(s) in display
Contact your Esco Medical distrib-
utor for PCB replacement
Absent or erratic function of keys
Contact your Esco Medical distrib-
utor to replace the keys