MAW
®
User Manual Rev. 3.0
49
27 Electromagnetic compatibility
Guidance and manufacturer’s declaration – electromagnetic emissions
The MAW, Multi-zone ART Workstation
®
is intended to be used in the electromagnetic environment specified
below. The customer or the user of MAW, Multi-zone ART Workstation
®
should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
MAW, Multi-zone ART Workstation
®
does not use RF energy.
Therefore, its RF emissions are very low and are not likely to
cause any interference with nearby electronic equipment
RF emissions
CISPR 11
Class A
MAW, Multi-zone ART Workstation
®
is suitable for use in a
hospital environment
It is not suited for domestic establishments
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
Class A
Guidance and manufacturer’s declaration – electromagnetic immunity
MAW, Multi-zone ART Workstation
®
is intended to be used in the electromagnetic environment specified be-
low. The customer or the user of MAW, Multi-zone ART Workstation
®
should assure that it is used in such an
environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic env
i
ronment-
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Flooring should be wood, con-
crete or ceramic tiles. If floor is
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast tran-
sient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/
output lines
Surge
IEC 61000-4-5
±1kV differential mode ±2kV
common mode
Voltage dips, short
interruptions and
voltage variations
on power supply in-
put lines
IEC 61000-4-11
<5 % 100V
(>95%dip in 100V) for 0.5 cycle
40% 100V (60% dip in 100V)
for 5 cycles
70% 100V (30% dip in 100V)
for 25 cycles) dip in 100V) for 5
sec
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Performance
A
The levels of power frequency
magnetic fields should be charac-
teristic to a typical location in a
typical commercial or hospital
environment.