Guidance and manufacturer’s declaration – electromagnetic immunity
MAW, Multi-zone ART Workstation
®
is intended to be used in the electromagnetic environment specified be-
low. The customer or the user of MAW, Multi-zone ART Workstation
®
should assure that it is used in such an
environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic env
i
ronment-
guidance
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
10 Vrms 150kHz to 80 MHz in
ISM bands
3V/m 80 MHz to 2.5 GHz
3V/m from
80MHz to 2.5
GHz
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of
MAW, Multi-zone ART Work-
station
®
, including cables, than
the recommended separation dis-
tance calculated according to the
equation, applicable to the trans-
mitter’s frequency
Recommended separation dis-
tance
d=0.35 P
d=0.35 P 80MHz to 800 MHz
d=0.7 P 800MHz to 2.5 GHz
Where P is the maximum
power output rating of the trans-
mitter in watts (W) according to
the transmitter manufacturer
and d is the recommended sepa-
ration distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be lower than the compli-
ance level in each frequency
range
Interference may occur in the vi-
cinity of equipment.