EasyManuals Logo

ESPEC Platinous ESL-3C Series User Manual

Default Icon
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
Troubleshooting
94
Platinous User’s Manual
Displayed Alarm
Code/Channel
Code (option)
Trouble Cause Remedial Action
COMMUNICATION
ERROR-5
(99)
The instrumentation
detected system trouble
in the CPU board. The
chamber has been
stopped.
Refrigeration task
error on CPU
board.
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.
COMMUNICATION
ERROR-6
(99)
The instrumentation
detected system trouble
in the CPU board. The
chamber has been
stopped.
Temperature
control task error
on the CPU
board.
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ESPEC Platinous ESL-3C Series and is the answer not in the manual?

ESPEC Platinous ESL-3C Series Specifications

General IconGeneral
BrandESPEC
ModelPlatinous ESL-3C Series
CategoryClimate chamber
LanguageEnglish