Control Solution Trouble Shooting
If your control solution testing is out of range (too high or too low), it
may be caused by the following:
Possible Causes What you can do…..
EvenCareG2
Control Solution Testing
29
Wrong brand of Control
Solution being used
Control solution not at room
temperature
Expired or contaminated
control solution or damaged
test strip.
Meter malfunction
Make sure you are using
EvenCare G2 Control Solution.
Make sure the testing environ-
ment is between 50˚F to 104˚F.
Check the expiration and open
bottle date on both the control
solution and test strips. Repeat
testing using a new test strip.
If the result is still out of range,
use a new bottle of control
solution and retest.
If the problem persists, please
call Medline’s Technical Service
Center at 1-866-274-2131
Monday through Friday between
8:00 a.m. and 5:00 p.m.(Central
Time). Contact your Healthcare
Provider at all other times.