Important Information on Alternate Site Testing (AST)
The EvenCare G2 Blood Glucose
Monitoring System can test for blood
glucose from areas other then your
fingertip such as palm and forearm
(alternate site testing, or AST).
Alternate site testing can be less
painful then fingertip testing, but
because of the physiological differ-
ences between your fingertip and
palm and forearm
1
; alternate site test
results maybe significantly different
then results from fingertip testing under certain conditions.
You should consult with your doctor or healthcare professional before
using alternate site testing.
DO AST ONLY in the following intervals:
• In a pre-meal or fasting state (more than 2 hours since the last meal)
• 2 hours or more after taking insulin.
• 2 hours or more after exercise.
Alternate Site Testing SHOULD NOT be used when:
You are Hypoglycemic unaware (not able to tell if you have
low blood sugar)
Within 2 hours of a meal, exercise, or medication
You will be operating machinery or driving a car.
During times of stress.
You are sick
You think your blood glucose is low.
Your AST results do not match the way you feel.
You are testing for hyperglycemia.
Your routine glucose results are often fluctuating.
EvenCareG2