Eversense CGM User Guide
Alerts and Notifications Displayed on the Apple Watch (continued)
Apple Watch
Display
Description
Sensor Suspend
Appears when the system’s internal checks
detect a need to restart the Initialization Phase
for additional calibrations. Glucose values will
be displayed a few minutes after the second
successful calibration during the Initialization
Phase.
Calibrate Now
Appears every 60 minutes when it is time for you
to calibrate.
Actions
Confirm the alert on your Eversense CGM
Mobile App. Re-initialization of the system
begins in 6 hours.
Confirm the alert on your Eversense CGM
Mobile App. Do a fingerstick blood glucose test
and enter the reading as your calibration value in
the mobile app. DO NOT use an alternative site
(such as forearm) to obtain your blood glucose
reading.
Alerts