12
04012XXX
ZSME400000011
EXCELL PRECISION CO., LTD
Item
Function
Setting value
Default
Parameter
Description
FNC-06
Front panel indication “◄
setting (top)
Parameter Description
0 Zero
1 Stable
2 Gross
3 Net
4 Totalised weight (Accu. V)
5 Totalised transactions (Accu. C)
6 SP1
7 SP2
8 SP3
9 Hi
10 OK
11 Lo
12 Under
13 Over
14 Discharge
15 Running
16 Hold
17 MD
0
FNC-07
Front panel indication “◄
setting (next to top)
1
FNC-08
Front panel indication “◄
setting (next to bottom)
2
FNC-09
Front panel indication “◄
setting (bottom)
3
FNC-10
Return to zero band
(d: refer to CSP-03)
0
5 d
0
1
10 d
2
20 d
3
40 d
4
60 d
5
80 d
6
100 d
7
150 d
8
200 d
9
250 d
FNC-11
Hold
0
Hold
0
1
Peak hold (positive 1)
2
Peak hold (negative)
3
Peak hold (absolute value)
4
Peak hold (positive 2)
FNC-12
Rate for display rewrite
0
No limitation
0
1
20 times/s
2
10 times/s
3
5 times/s
4
1 time/s

Table of Contents

Related product manuals