EasyManuals Logo

Excell 320S User Manual

Excell 320S
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
15
04012XXX
ZSME400000011
EXCELL PRECISION CO., LTD
2-4 Error Messages (Display in General Function setting)
. Load Cell output voltage - 0.1m V / V or 4mV / V
. Weight value previous weight value
. Actual measured weight value previous weight value
. Setting value 0
. mV / V value entered measuring range
. mV / V value entered is too small (SPAN Zero 0 mV / V)
. Displayed resolution is less than 0.12 V / division
2-5 Set internal calibration password (FNC-16)
The default value of FNC-16 is 0000, which means internal calibration without password. To
set a password, enter 4-digit numbers (other than 0001, 0000) and press confirm key to
complete. If password is set to 0001, error message UN.VALI will appear.
If password has been set, FNC-16 will display 0001. To change password, it must enter
the previously set password first. If entered password is wrong, it displays PW ERR. If entered
password is correct, it displays NEW PW and then enters 4-digit numbers (other than 0001) for
new password or enters 0000 to remove password. If password is forgotten, it must reset to
factory setting.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Excell 320S and is the answer not in the manual?

Excell 320S Specifications

General IconGeneral
BrandExcell
Model320S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals