Excell 320S - FNC Settings and MODBUS Test; Zero, Stand-by, and Password Settings; MODBUS TCP;IP Connectivity Test

Excell 320S
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
04012XXX
ZSME400000011
EXCELL PRECISION CO., LTD
Item
Function
Setting value
Default
Parameter
Description
FNC-13
Turn-on zero setting
0
Disable
0
1
Enable
FNC-14
Stand-by mode setting
0
Disable all the functions under
stand-by mode
0
1
Only turn off display but not
disable other functions under
stand-by mode
FNC-15
Zero function record
setting
0
Zero point record not saved into
EEPROM
0
1
Zero point record saved into
EEPROM
FNC-16
Internal calibration
password setting
0000
No password
0000
0001
Password is set
FNC-17
Turn on/off watchdog
1
Turn on watchdog
1
0
Turn off watchdog
FNC-18
China Type Approval
0
Non-approval
0
1
China Type Approval
2-3 MODBUS Over TCP/IP Test Method
1. Connect 320S and PC/Notebook to the same HUB with RJ-45 cable.
2. Turn on 320S
3. Open terminal or PowerShell software on PC/Notebook side and execute ping
192.168.1.210
If the 320S does not respond, please recheck the network connection or restart the 320S