Page. 37
10. Service parts䯂Option parts
Ɣ Call your local Fellow Kogyo distributor or Sales Division of Fellow Kogyo when you want to purchase maintenance parts or components.
䦟
䎃 Service parts (accessories)
Part name Part No. Qty.
Rating/
specification
Part name Part No. Qty.
Rating/
specification
Fuse for the equipment 2000-010
1 pc
3.15A, 250V
I10 joint set 2000-050
3 pcs
RC1/4䐙1
RC1/8䐙2
Fuse for the valve
Electromagnetic valve
unit
2000-020
1 set
(2 pcs)
2A, 250V
Upper lid O-ring 2000-060
1 pc
4DP-31
Washing brush 2000-030
1 pc
Drain seal W 2000-070
1 pc
M5
Washing liquid 2000-040
1 pc
125mL
䦟䎃 Option parts
Part name Part No. Qty.
Rating/
specification
Part name Part No. Qty.
Rating/
specification
Sensor unit
2000-210
1 pc Calibrated Upper lid
2000-260
1 pc
Electromagnetic valve unit
2000-220
1 pc
Lithium battery
2000-270
1 pc
CR2450
Behavior verification kit
2000-230
1 set
Fresh water filter set 2000-280
1 set
Desiccant (silica gel)
2000-240
1 set
Sample water filter set 2000-290
1 set
Operational log memory
card
2000-250
1 pc
Ɣ When effecting every-5-years renewal of precision certificate, you can replace calibrated sensor unit (sold separately) on board and use it.
Ɣ You can check the operation (e.g., PSC inspection) of the oil concentration detector by the Operation Test Kit (separately sold).
(The accuracy will not be assured.)
11.䎃 Request
The FOCAS-2000 oil concentration detector is a high precision instrument capable of detecting oil in parts per million. To prevent the product
against failures and maintain its accuracy, the product must be calibrated within 5 years by the manufacturer or the agent authorized by the
manufacturer and have the IOPP certificate renewal examination.
(It is permitted to replace your product by a calibrated and verified product.)
Ɣ For ensuring precision, it is recommended that the user have its equipment calibrated or verified for precision by the agent or
manufacturer every one to two years.
Ɣ Checkup / repair or precision verification after shipment must be performed in accordance with IMO regulations by a person who has
skills certified by Fellow Kogyo.
Ɣ Breaking the seal inside the equipment is deemed as a fraudulent act. Never break the seal.
Ɣ Do not access any portion except for portions specified in this operation manual. Access to unspecified portions is prohibited by law.
Contamination is minimized through keeping the detection tube clean by automatically performing washing and filling the detection portion
with fresh water after using the equipment.
Ɣ Ensure to install a fresh water line. Fresh water for use is recommended to be equivalent to mineral water.
Ɣ Care must be taken if the equipment is used in a place with the possibility of freezing. Drain water completely from the detective tube
by using the manual drain function if the equipment is stored in the environment of low temperatures or left with the detection tube
water-filled for a prolonged time.