Page. 6
2. Specifications
■ 2-4. Relay contact output, alarm lamp movement view
Terminal name /
operation conditions etc
Power supply source
Without
supply
COM-NC
During
power
outage
Operational
status
BLG-NA
NO-COM
Alarm1 lamp (AL1)
between relay terminal
※3
Alarm2 lamp (AL2)
Default value
(At factory shipment)
Arbitrary setting
Unit
(mA)
Arbitrary setting is possible via panel switch operation
See Item 6-3-1 “Setting data and ranges” under Section 6 “Setting”
※During power outage and Warm up is in-possible to arbitrary setting.
Alarm1
Alarm signal 1
Warm up
Under set
density valve
Over30ppm
Over scale
During
setting
Defect error
More than set
density value
RECout
(output of 4 to 20 mA)
OFF
OFF
ON
SettingWashing
OFF
ON
OFF
NA-SEA
COM-NC
NO-COM
COM-NC
between relay terminal
2
VALVE for
theree-way
valve
Across the relay terminals
(Applied voltage output)
Across the relay terminals
(Applied voltage output)
Specification A
Specification B
※1
※3
Alarm2
Alarm signal 2
0
(No output)
Standby
Standby During operation
With supply
0
OFF
ON
OFF
OFF
ON
OFF
ON
(Blinking)
4
※2
※2
OFF
OFF
ON (Voltage output)
OFF
ON
(Voltage output)
ON
(Voltage output)
OFF
ON
OFF
OFF
ON
OFF
※4
Concentra-
tion monitor
※4
20 4 0
0
OFF
ON
OFF
ON
※2
※2
OFF
ON
OFF
ON
ON(Blinking)
OFF
OFF
ON (Voltage output)
3
5
4
1
Concentra-
tion monitor
ON(Blinking)
ON
(Blinking)
※1 Terminal bases are different. (See "2-3-2 "Terminal Base Alignment and FOCAS-2000 inner connection.)
※2 You can set to synchronize alarm signals output from contacts with alarm lamps by operating the panel switch (when you set, wash, and warm up the detector).
Initially (when the detector is shipped from the factory), its panel is set to output no contact alarm but the alarm lamps singly blink (to alarm).
※3 Alarm1 and Alarm2 values can be set individually (in ppm) in the range of 5 to 15 ppm.
※4 RECout (4-20mA output) outputs a direct current of 4 to 20mA in synchronism with the measured oil concentration value.