Moisture Measurement Operation
7
Moisture-only mode: The reference value, with delta symbol, is shown on
the right side of the display and the moisture reading (offset by the refer-
ence value) is shown on the left in large digits.
1. All measurements taken subsequently will be relative to the reference val-
ue. For example, if the reference value is ‘20’ (representing the driest area
of the material under test) and a measurement of ‘25’ is taken (in an area
with higher moisture content), the measurement will show ‘5’ (25 20 = 5).
As implied, this mode is useful for comparing wet areas to a dry area
reference.
2. To remove the reference value and exit the mode: Remove the meter sen-
sor from the area under test, so that the sensor is no longer touching a
surface and is clear of any objects, and then select the pin-mode from the
Moisture Mode menu.
7.9 High Moisture Alarm
1. To access the Alarm mode use the menus (Settings/Measurement Param-
eters/IR Temperature & Moisture Alarm). See Section 8.4 for High/Low
Temperature Alarm mode.
Figure 7.14 Setting the Moisture Alarm
2. Select the ‘Set Moisture Alarm’ mode.
3. Use the navigation arrows and Select button
to set the Alarm to
ABOVE’ or ‘OFF’, to set the max. threshold (0% to 100%), and to select
‘No Sound’ or ‘Beep’ for the alarm alert; the visual alert is always active.
Press Select to confirm and press Return
to exit the menu.
4. When the High Alarm is enabled, the main display will show the alarm bell
icon next to the measurement digits. Refer to Section 5.8 Settings menu
for additional information.
#NAS100005; r. AL/76939/76957; en-US
38