EasyManua.ls Logo

Flomec RT40 - Calibration Data

Flomec RT40
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
5.7 Calibration Data
The calibration data menu allows the user to enter up to 5 points of calibration data for their
flow meter; if the RT40 was purchased integrally mounted to a flow meter then the calibration
data will be factory set.
Upon entering the calibration data menu the user is able to enter up to 5 points of
calibration, with each point of calibration requiring a frequency value and a K-factor.
Frequency values are to be entered in Hertz (pulses/second) and can be entered with up to 4
whole numbers and 1 decimal place. K-factor values are to be entered in Pulses/Litre with up
to 6 whole numbers and 4 decimal places.
It is very important to note that K-factors are always entered in units of Litres regardless of
the units chosen for the display.
If a single point of calibration is to be entered, the frequency value is not important and any
non-zero value can be entered here. After entering a non-zero value for frequency the K-factor
for the flowmeter may be entered. Please note, if there are no non-zero frequency values
entered the RT40 will operate as if the K-factor has been set to 1 Pulses/Litre.
When entering multiple points of calibration the user must calculate the frequency for each
calibration point, which can be done as follows:



. 


÷ 60 = ()
Example: for a calibration point at 50 L/min with a K-factor of 107 Pulses/Litre the user
would enter a frequency value of 89.2Hz and a K-factor of 107 Pulses/Litre
Calibration points may be entered in any order, it is not necessary to enter them in ascending
frequency order. However, if any point is entered with a frequency value of zero then all
remaining calibration points will be disabled. If data is entered for the first 3 points of
calibration, and point 4 is entered as zero Hz then points 4 and 5 will be disabled and the RT40
will operate with 3 points of calibration.
For users with a QS200 or QS100 ultrasonic water meter; the calibration data to enter should
be a frequency of 1Hz, and a K-factor selected from the “Reference Pulses/Litre” column of
the K-factor table in the QS200 or QS100 instruction manual. Please note that the correct
Pulses/Litre number must be selected based on your pipe size and your pipe schedule.

Related product manuals