6.2 Troubleshooting Summary
Symptom Probable Cause Corrective Action
INACCURATE
READINGS
1. Incorrectly entered
calibration data
2. Electrical interference
3. Mechanical or electrical
fault with flowmeter
4. Flowmeter requires
recalibration
See 6.3.2 below
Contact the manufacturer if a
product fault is identified
TOTAL NOT
COUNTING WITH
FLOW IN PIPE
1. Incorrect configuration
2. Incorrect wiring
3. Product fault
See 6.3.1 below
Contact the manufacturer if a
product fault is identified
ZERO FLOWRATE
DISPLAY WITH
FLOW IN PIPE
1. Incorrect configuration
2. Product fault
See 6.3.2 below
Contact the manufacturer if a
product fault is identified
UNEXPECTED
CHARACTERS ON
LCD SCREEN
Faulty LCD segment
Enter programming mode to test
display segments. Confirm that
all LCD segments are illuminating
as per image on page 6 of this
instruction manual. If a
segment(s) is not illuminating
contact the manufacturer
NO PULSE
OUTPUT
RECEIVED
1. Incorrect configuration
2. Incorrect wiring
3. Faulty receiving instrument
(PLC, FMS, etc.)
4. Product fault
See 6.3.5 below
Contact the manufacturer if a
product fault is identified