EasyManua.ls Logo

Flowserve APEX 7000 - Appendix A: Calibration; APEX 7000 Calibration Guide

Flowserve APEX 7000
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EP5 Electro Pneumatic Digital Positioner FCD PMENIM0006-00-A5 12/18
33
Calibration Procedure APEX 7000:
1. Apply 0% input signal (0% = 20 kPa, 3 psi, or 4 mA).
2. Wait for steady state. It is important to wait for steady state. On very large actuators, it can take minutes to
establish.
3. Loosen locking screw A about 1 turn, if it is looser, you risk introducing unwanted movement.
4. Adjust “ZERO” position using thumbwheel B
5. Apply 100% input signal (100% = 100 kPa, 15 psi, or 20 mA).
6. Wait for steady state.
7. With a screw drive, adjust the “SPAN” setting by turning screw C, while doing so, keep the zero screw B in
position by applying pressure with your thumb.
8. Lock the calibration mechanism in place by tightening locking screw A
9. Apply 0% signal and verify zero position.
Apendix A

Related product manuals