35
Display Functions and User Interface
RECALL
4.8 ENTER
The ENTER key is used to select a menu option, accept a setting change, congura-
tion change, or enable a logging session.
4.9 RECALL
The 1523/24 has a Demand Log that stores up to 25 readings that may be displayed on
screen or sent out the serial port to a computer. The RECALL menu allows the user to
review, delete, or send demand log data out the serial port.
The 1524 has an Auto-Log with the added capability of storing up to 15,000 readings
with a time and date stamp on each reading. The readings can be sent to a computer or
deleted using this menu.
4.9.1 Review Saved
This menu displays one of up to 25 saved readings on main screen for review. The
display shows channel, auxiliary display value, main display value, tag id and time
stamp if model 1524. The user can scroll through each log by pressing the up and
down arrow keys. The user can exit this review mode by pressing the RECALL or the
HOME (SHIFT + ENTER) keys.
4.9.2 Send Saved
This feature sends the entire demand log out the serial port to a computer. When all of
the data is sent, the display will show “100%” to the right of the Send Saved command
selection to indicate that the send function is complete. This does not affect the stored
data. Refer to the Demand Log, and Digital Communications Interface section of this
manual for more information.
4.9.3 Delete Saved
This feature allows the user to delete data les saved in the Demand Log. The user
may delete a specied le or all les. Use the Up and Down arrow keys to select the
le to delete. The instrument default is used to delete all les. The user can exit this
menu and return to the main display by pressing the RECALL or HOME (SHIFT +
ENTER) key(s).
4.9.4 Send Logs
This feature allows the user to send logged data les out the serial port to a computer.
In order to use this feature the user must turn the serial port on, and set the baud rate
to the appropriate setting. Refer to Section 5.2, AUTO-LOG on page 39, and Section
6, Digital Communication Interface on page 43for more information. This feature is
available on model 1524 only.
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.com