1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
76
Table 23 Demand Log Statistical Types
Number Type
1 Maximum
2 Minimum
3 Average
4 Standard Deviation
5 Auxiliary Value
For all query commands, when an error is detected and an error message is placed in
the queue, no response is returned for the query.
For all action commands, when an error is detected and an error message is placed in
the queue, no action is taken.
If a thermocouple is inserted into channel 2, all commands except the SYST:ERR
query command place the error ‘-221, “Settings conict”’ in the queue. This action
complements the full lockout of all the keypad keys in this same circumstance.
Error messages are retrieved from the queue using the SYST:ERR query.
Table 24 Error Messages
Number Message
0 "No error"
-100 "Command error"
-200 "Execution error"
-203 "Command protected"
-221 “Settings conict”
-350 "Queue overow"
-360 "Communication error"
-363 "Input buffer overrun"
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1523 and is the answer not in the manual?

Fluke 1523 Specifications

General IconGeneral
Resolution0.001°C
DisplayLCD
Data LoggingYes
Input TypesRTD, Thermocouple
InterfaceRS-232
Power SupplyAC adapter
Accuracy±0.005°C

Related product manuals