79
Calibration of Your Reference Thermometer Readout
Manual Calibration
Figure 18
Prepare Test
Equipment
Visual
Inspection
Need As-
Found data?
As-Found
Data
Parameters
Yes
As-Found Test
Data
Set Reference
Points and Reset
Adj Values
No
Take Alignment
Data
Calculate New Adj
Values
As Left Accuracy
Data
Finished
Flow chart for manual calibration