1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
58
Error ‘-203,”Command protected”’ is queued if the password has not been entered
with SYST:PASS:CEN.
CALibrate<chn>:DEVice:RANGe:SELection?
Return the range presently being calibrated for the specied channel.
The CALibrate sufx, <chn>, species the channel, 1 or 2. Defaults to 1 if omitted.
The response is one of the keywords listed in Table 22 on page 75.
Example: CAL:DEV:RANG:SEL?
Response: LO_OHMS
CALibrate<chn>:DEVice:RANGe:SELection <val>
Set the range presently being calibrated for the specied channel.
The CALibrate sufx, <chn>, species the channel, 1 or 2. Defaults to 1 if omitted.
The parameter <val> is one of the keywords in Table 22 on page 75. The value NOR-
MAL exits calibration mode, reconnects the inserted probe if any, and refreshes the
display.
This command sets the range to which all of the following commands apply:
CAL:DEV:RANG:CLEA
CAL:DEV:RANG:NUM
CAL:DEV:RANG:REF
CAL:DEV:RANG:ADJ
Example: CAL:DEV:RANG:SEL LO_OHMS
DISPlay<chn>:RESolution?
Return the maximum display resolution setting for the channel.
The DISPlay sufx, <chn>, species the channel, 1 or 2. Defaults to 1 if omitted.
Example: DISP1:RES?
Response: 2
DISPlay<chn>:RESolution <num>
Set the maximum display resolution for the channel.
The DISPlay sufx, <chn>, species the channel, 1 or 2. Defaults to 1 if omitted.