EasyManuals Logo

Fluke 1523 User Manual

Fluke 1523
105 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
1523, 1524 Reference Thermometer
Serial Commands - Alphabetic Listing
72
Example: SYST:PASS:NEW 4321
This command has no response.
NOTE: This command is protected, which requires a password to set it.
Error ‘-203,”Command protected”’ is queued if the old password has not been entered
with SYST:PASS:CEN.
SYSTem:SNUMber?
Return the device serial number.
If the serial number has not been set at the factory, a 0 is returned.
Example: SYST:SNUM?
Response: 509002
SYSTem:TIME?
Return the current system time in 24 hour format.
The response is returned in the 24 hour format, <hour>,<minute>,<second>.
This command is only available in the model 1524, and results in a command error if
used with the model1523.
Example: SYST:TIME?
Response: 18,43,23
SYSTem:TIME <hour>,<minute>,<second>
Set the current system time.
The <hour>, <minute>, and <second> parameters must all be supplied.
The <hour> parameter is a one or two-digit number, 0 to 23.
The <minute> parameter is a one or two-digit number, 0 to 59.
The <second> parameter is a one or two-digit number 0 to 59.
This command is only available in the model 1524, and results in a command error if
used with the model1523.
Example: SYST:TIME 18,43,23
SYSTem:VERSion?
Return the SCPI version number.
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1523 and is the answer not in the manual?

Fluke 1523 Specifications

General IconGeneral
Resolution0.001°C
DisplayLCD
Data LoggingYes
Input TypesRTD, Thermocouple
InterfaceRS-232
Power SupplyAC adapter
Accuracy±0.005°C

Related product manuals