Insulation Tester
The Tester
9
Also use
L M to access these menu items:
Voltage adjustment in 100 V for each step
Time limit xx-xx
T - 0 to 99 minutes
T1, T2, T3 - T1 < T2 < T3
T1 > 0 seconds, T3 <1000 seconds
Show test parameters
Change the name of the ID tag for the test result
•Push
I to make the selection.
Te s te r O n /O f f
Push A to turn on the Tester.
The Tester does a self-check, shows the software version, and starts in the
Insulation Resistance Idle mode.
In Insulation Resistance Idle mode:
Change test parameters
Start an insulation test
View stored test results (1537 only)
Download test results (1537 only)
W Caution
If the object under test is highly capacitive, it can take a long time
for the Product to discharge. Do not turn off the Product or remove
test leads until the discharge is complete.
When on, push
A >1 second to turn off the Tester.