Insulation Tester
V ac / V dc / Resistance (1537 only)
23
Delete Test Results Stored in Memory (1537 only)
You can delete a selected test result or delete all stored test results.
To delete a selected test result:
1. Push I to view stored records.
2. Push
J K to select Record.
3. With the record selected, push
H. The displays shows the blinking
message: DELETE?
4. Push
I to delete the current record or push G to cancel.
To delete all test results:
1. Push
I to view stored records.
2. Push
H >2 seconds. The displays shows the blinking message: DELETE
ALL?
3. Push
I to delete all records or push G to cancel.
V ac / V dc / Resistance (1537 only)
The 1537 includes DMM functions for V ac / V dc and resistance
measurements.
To do a V ac / V dc or resistance test:
1. Turn on the Tester.
2. Push F to select the V ac / V dc or resistance function.
3. Insert the test leads into the correct terminals. See Ta b l e 8 .
4. Connect the test leads to the circuit under test.
The test result appears on the Tester when the test is complete.
Note
The Tester does not support test result records for the V ac / V dc /
Resistance measurements.
The Tester alarm beeps if the resistance measurement is 30 Ω.