EasyManua.ls Logo

Fluke 1537-II - Page 34

Fluke 1537-II
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
Insulation Tester
Electrical Specications
Table 12. V ac / V dc / Resistance Measurement (1537/1537-II Only)
Function Range Resolution
Accuracy
±(% reading + Digits)
V ac 0 V to 600.0 V 0.1 V
±(2 % +10)
(45 Hz to 500 Hz)
V dc 0 V to 600.0 V 0.1 V ±(2 % +10)
Resistance
0 Ω to 600.0 Ω 0.1 Ω
±(2 % +10)600 Ω to 6000 Ω 1 Ω
6.00 kΩ to 60.00 kΩ 0.01 kΩ
Table 13. Operating Ranges and Uncertainties per EN 61557
Function Display Range
EN 61557
Measurement
Range Operating
Uncertainty
±(% reading + Digits)
Nominal Values
V
EN 61557- 1
[1]
0 V ac to 600 V ac
45 Hz to 500 Hz
0 V ac to 600 V ac
±(2 % +10 digits)
UN = 230/400 V ac
f = 50/60 Hz
RISO
EN 61557-2
0 kΩ to 500 GΩ
200 kΩ to 500 GΩ
±20 %
UN = 250 / 500 /
1000 / 2500 V dc
IN = 1.0 mA
[1]
1537/1537-II only
Table 14. Operating Uncertainties per EN 61557
Parameter Specication Typical Max
[1]
Intrinsic
uncertainty
IEC 61557-2
A, Reference conditions
1.63 % 3.68 %
Position
IEC 61557-2
E1, Reference Conditions ±90 °
2.29 % 5.00 %
Supply Voltage
IEC 61557-2
E2, At low battery voltage
stated by manufacture
2.80 % 6.09 %
Temperature
IEC 61557-2
E3, -10 °C and 50 °C
3.36 % 9.83 %
Operating
Uncertainty
IEC 61557-2
B, ≤30 %
7.30 % 18.17 %
[1]
Condence level: 95 %