1652C/1653B/1654B
Calibration Manual
30
Table 7. RCD Trip Current Accuracy Tests
UUT Settings Current Accuracy
Step
UUT
Function
<
Spec
Lo Limit Hi Limit
1 10 mA x1/2 AC 0° -10 % + 0 4.50 5.00
2 10 mA x1/2 A 0° -10 % + 0 3.15 3.50
3 10 mA x5 AC 0° +10 % - 0 50.0 55.0
4 30 mA x1 AC 0° +10 % - 0 30.0 33.0
5 30 mA x1 AC 180° +10 % - 0 30.0 33.0
6 30 mA x1 A 0° +10 % - 0 42.4 46.7
7 30 mA x1/2 AC 0° -10 % + 0 13.5 15.0
8
[1]
30 mA x1 B 0° +10 % - 0 60 66
9
[1]
30 mA x1 B 180° +10 % - 0 60 66
10
[1]
300 mA x1 B 0° +10 % - 0 600 660
11
T
1000 mA x1 AC 0° +10 % - 0 1000 1100
[1] Steps 8-10 = 1654B only.