EasyManuals Logo

Fluke 1736/FPC Calibration Manual

Fluke 1736/FPC
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Logger/Monitor
Accuracy Verification Procedure
21
Accuracy Verification Procedure
The procedure verifies the Power Logger accuracy at ambient temperature 23 °C ±2 °C (intrinsic error).
A complete accuracy verification of the Fluke 173x consists of:
Voltage Measurement
Current Measurement
AUX Measurement
Optional Flexi or Current Clamp Verification
Voltage Measurement
1. Select the setup. See Basic Instrument Setup for all Verifications.
2. Connect the VL1730 "N" lead to the calibrator NORMAL LO.
3. The Logger must be on battery power with 50 % charge.
4. Connect the calibrator NORMAL V output to the VL1730 L1+L2+L3 leads.
5. Sequentially set the calibrator to the voltages indicated in Table 10 and check that the Logger reading is
between the limits.
6. Do this for all ranges indicated in Table 10:
set the calibrator to supply a 57.0 Hz sine wave for all voltages
wait until each reading has stabilized
The spreadsheet is the first choice for readings. Readings will have more resolution from the
spreadsheet.
7. Push to select the Power Logger voltage display.
Table 10. Voltage Verification
8. When you are done, set the calibrator to Standby.
Nominal Voltage
(Range)
Calibrator voltage
(57 Hz sine wave)
Minimum Reading
± (0.2 % + 0.01 %)
Maximum Reading
1000 10 9.9 10.1
1000 100 99.7 100.3
1000 500 498.9 501.1
1000 1000 998 1002

Other manuals for Fluke 1736/FPC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1736/FPC and is the answer not in the manual?

Fluke 1736/FPC Specifications

General IconGeneral
BrandFluke
Model1736/FPC
CategoryData Loggers
LanguageEnglish

Related product manuals