True-rms Remote Display Digital Multimeter
How to Make Measurements
19
Temperature Measurements
The Meter measures the temperature of a type-K
thermocouple (included). Choose between degrees
Celsius (°C) or degrees Fahrenheit (°F) by pushing
C.
W Caution
To prevent possible damage to the Meter or
other equipment, use a thermocouple rated
for the temperatures to be measured. The
Meter is rated for -40.0 °C to +400.0 °C and
-40.0 °F to 752 °F, but the included type-K
thermocouple is rated to 260 °C.
The temperature ranges are -40.0 °C to +400 °C and
-40.0 °F to 752 °F. All other temperatures show  on the
display. When there is no thermocouple connected, the
display shows .
To measure temperature:
1. Connect a type-K thermocouple to the COM and
terminals of the Meter.
2. Turn the function switch to .
3. Push C to choose Celsius or Fahrenheit.
Continuity Tests
WCaution
To prevent possible damage to the Meter or
the equipment under test, disconnect power
and discharge all high-voltage capacitors
before a continuity test.
The continuity test uses a beeper that sounds when a
closed circuit is sensed. The beeper lets you do continuity
tests without the necessity to look at the display.
To do a continuity test, set up the Meter as shown in
Figure 3.