True-rms Remote Display Digital Multimeter
Detailed Specifications
39
Weight ............................................................................604 g (1.3 lbs)
Safety Compliance ........................................................Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 61010-1-04 to 1000 V
Measurement Category III and 600 V Measurement Category IV.
Certifications .................................................................CSA, TÜV (EN61010), P, ; (N10140),VDE, GOST
Detailed Specifications
For all detailed specifications:
Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy
specifications take the form of ±([ % of Reading ] + [ Number of least significant digits ]).
AC Voltage
AC conversions are ac-coupled and valid from 1 % to 100 % of range.
Accuracy
Range
[1]
Resolution
45 – 500 Hz 500 Hz – 1 kHz
600.0 mV 0.1 mV
6.000 V 0.001 V
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V
±(1.0 % + 3) ±(2.0 % + 3)
[1] Crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale.