EasyManua.ls Logo

Fluke 233 - Current, Capacitance, Diode Specs

Fluke 233
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
True-rms Remote Display Digital Multimeter
Detailed Specifications
41
AC Current
Function Range Resolution
Accuracy
(45 – 500 Hz)
6.000 A 0.001 A
A ac
[1,2,3]
10.00 A 0.01 A
±(1.5 % + 3)
[1] All ranges are specified from 5 % of range to 100 % of range.
[2] Crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale.
[3] AC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum.
DC Current
Function Range Resolution Accuracy
6.000 A 0.001 A
A dc
[1]
10.00 A 0.01 A
±(1.0 % + 3)
[1] DC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum.

Other manuals for Fluke 233