27 II/28 II 
Users Manual 
16 
When measuring voltage, the Meter acts approximately 
like a 10-MΩ (10,000,000 Ω) impedance in parallel with 
the circuit. This loading effect can cause measurement 
errors in high-impedance circuits. In most cases, the error 
is negligible (0.1 % or less) if the circuit impedance is 
10 kΩ (10,000 Ω) or less. 
For better accuracy when measuring the dc offset of an 
ac voltage, measure the ac voltage first. Note the ac 
voltage range, then manually select a dc voltage range 
equal to or higher than the ac range. This procedure 
improves the accuracy of the dc measurement by 
ensuring that the input protection circuits are not 
activated. 
Zero Input Behavior of True-rms Meters (28 II) 
True-rms meters accurately measure distorted 
waveforms, but when the input leads are shorted together 
in the ac functions, the Meter displays a residual reading 
between 1 and 30 counts. When the test leads are open, 
the display readings may fluctuate due to interference. 
These offset readings are normal. They do not affect the 
Meter’s ac measurement accuracy over the specified 
measurement ranges. 
Unspecified input levels are: 
•  AC voltage: below 3 % of 600 mV ac, or 18 mV ac 
•  AC current: below 3 % of 60 mA ac, or 1.8 mA ac 
•  AC current: below 3 % of 600 μA ac, or 18 μA ac 
Low-Pass Filter (28 II) 
The 28 II is equipped with an ac low-pass filter. When 
measuring ac voltage or ac frequency, press  to 
activate the low-pass filter mode (). The Meter 
continues measuring in the chosen mode, but now the 
signal diverts through a filter that blocks unwanted 
voltages above 1 kHz, refer to Figure 3. The lower 
frequency voltages pass with reduced accuracy to the 
measurement below 1 kHz. The low-pass filter can 
improve measurement performance on composite sine 
waves that are typically generated by inverters and 
variable frequency motor drives.