27 II/28 II
Users Manual
50
DC Current
Accuracy
Function Range Resolution Burden Voltage
27 II 28 II
600.0 μA 0.1 μA 100 μV/μA ±(0.2 % + 4) ±(0.2 % + 4)
μA dc
6000 μA 1 μA 100 μV/μA ±(0.2 % + 2) ±(0.2 % + 2)
60.00 mA 0.01 mA 1.8 mV/mA ±(0.2 % + 4) ±(0.2 % + 4)
mA dc
400.0 mA
[1]
0.1 mA 1.8 mV/mA ±(0.2 % + 2) ±(0.2 % + 2)
6.000 A 0.001 A 0.03 V/A ±(0.2 % + 4) ±(0.2 % + 4)
A dc
10.00 A
[2,3]
0.01 A 0.03 V/A ±(0.2 % + 2) ±(0.2 % + 2)
[1] 400 mA continuous; 600 mA for 18 hr maximum.
[2] W 10 A continuous up to 35 °C. <20 minutes on, 5 minutes off at 35 °C to 55 °C. >10 A to 20 A for 30 seconds maximum, 5
minutes off.
[3] >10 A accuracy unspecified.
Capacitance
Range Resolution Accuracy
10.00 nF 0.01 nF
100.0 nF 0.1 nF
±(1.0 % + 2)
[1]
1.000 μF 0.001 μF
10.00 μF 0.01 μF
100.0 μF 0.1 μF
9999 μF 1 μF
±(1.0 % + 2)
[1] With a film capacitor or better, using the rel mode to zero residual.