Digital Multimeters
Detailed Specifications
53
MIN MAX Recording
Accuracy Nominal Response
27 II 28 II
100 ms to 80 %
Specified accuracy ±12 counts for changes
>200 ms in duration
(±40 counts in ac with beeper on)
100 ms to 80 %
(dc functions)
Specified accuracy ±12 counts for
changes >200 ms in duration
120 ms to 80 %
(ac functions)
Specified accuracy ±40 counts for
changes >350 ms and inputs >25 %
of range
250 μs (peak)
[1]
Specified accuracy ±100 counts for
changes >250 μs in duration
(add ±100 counts for readings over
6000 counts)
(add ±100 counts for readings in Low
Pass mode)
[1] For repetitive peaks: 1 ms for single events.