27 II/28 II
Users Manual
46
Detailed Specifications
For all detailed specifications:
Accuracy is specified for 2 years after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 95 %.
Accuracy specifications take the form of ±([% of Reading] + [Number of least-significant digits]). For Model 28 II in the 4 ½-digit mode,
multiply the number of least-significant digits (counts) by 10.
27 II AC Voltage
Accuracy
[2]
Range Resolution
40 Hz – 2 kHz 2 kHz – 10 kHz 10 kHz – 30 kHz
600.0 mV 0.1 mV
6.000 V 0.001 V
60.00 V 0.01 V
±(4 % + 10)
600.0 V 0.1 V
±(0.5 % + 3) ±(2 % + 3)
±(4 % + 10)
[1]
1000 V 1 V ±(1.0 % + 3) ±(3 % + 3) Unspecified
[1] Specified to a maximum of 300 V ac
[2] Below 5 % of range, temperature coefficient is 0.15 X (specified accuracy)/ °C (>40 °C).