Digital Multimeters
General Specifications
45
Electromagnetic Compatibility (EN 61326-1:1997).....In an RF field of 3 V/M, accuracy = specified accuracy +20 counts, except 600 μA dc
range total accuracy = specified accuracy +60 counts. Temperature not specified
Relative Humidity ..........................................................0 % to 95 % (0 °C to 35 °C)
0 % to 70 % (35 °C to 55 °C)
Battery Type...................................................................3 AA Alkaline batteries, NEDA 15A IEC LR6, MSHA approved for use with three
Energizer P/N E91 or three Duracell P/N MN1500 1.5 Volt, AA alkaline batteries
only.
Battery Life.....................................................................800 hrs typical without backlight (Alkaline)
Vibration.........................................................................Per MIL-PRF-28800 for a Class 2 instrument
Shock..............................................................................1 Meter drop per IEC 61010 (3 Meter drop with holster)
Size (H x W x L)..............................................................1.80 in x 3.95 in x 8.40 in (4.57 cm x 10.0 cm x 21.33 cm)
Size with Holster............................................................2.50 in x 3.95 in x 7.80 in (6.35 cm x 10.0 cm x 19.81 cm)
Weight ............................................................................1.14 lb (517.1 g)
Weight with Holster and Flex-Stand ............................1.54 lb (698.5 g)
Safety Compliance ........................................................Complies with ANSI/ISA S82.01-2004, CAN/CSA C22.2 61010-1-04 to 600 V
Measurement Category IV. Licensed by TÜV to EN61010-1
Certifications .................................................................CSA, TÜV, P,
;, GOST, s
IP Rating.........................................................................67 (Protected against dust and the effect of immersion between 15 cm and 1 m for
30 min)
MSHA Approval No .......................................................18-A100015-0