Using
Source Mode
NOTE
ifyou use
copper wire
instead of thermocouple wire,
the reference junction is
at
the input terminals
of the
instrument under test
NOT inside the
calibrator.
Measure
the
external reference temperature and enter it
as
follows:
Press
feH
.
Set
Ref. June.
Compensat. to External. Enter the external
reference value.
Simulating RTDs
1 . Connect the calibrator to the instrument under test as shown.
2.
IsourceI
for SOURCE mode.
3.
Press UnTl
twice.
4.
®
to select the desired RID type.
5.
6. Enter the desired temperature.
7. |
ENTER
I,
RTD
Transmitter
Stepping the
Output
With the step feature,
you can increment or decrement the output
by
pressing
the
® or
<$>
arrow keys.
You
can use the step featu re for sourcing
electrical parameters and temperatures,
but not pressure.