Calibrating
a Process
Instrument
7.
IsSuAy for
MEASURE/SOURCE mode. The
display
changes as
shown below:
4.001
mA
100.0
int.
Ref. 26.3 °C
C Type J
°c
3.328 mV
As
Found
KM
8.
As Found softkey.
The display changes as shown below:
0% Value
100% Value
Tcierance
User- Entered Value
MB
mA
mm
mA
?rm
%
Disabled
0% Value
1 00% Value
Test Strategy
TC Type
K
77777? °C
mm
x
mm
Abort
Done
9. The display prompts
you to enter values for
0%,
100% and Tolerance.
1 0. User Entered Value
is for a value measured
by some other device, such as
the reading
reported
by
the control room. Leave
this set to Disabled because
you
are
only interested in values
measured at the transmitter output.
1 1 .
Press
the
®
key to move
the cursor down to enter 0% and 100% values for
SOURCE temperature.
Our example
uses
100.0
and 300.0°C.
1
2. The Test Strategy is
the number of test steps
and
whether
the
test
steps
are
performed rising and
failing in percent
of scale, or just
rising
or just falling.
Our example
uses five steps
(0%,
25%,
50%, 75%,
and 1
00%),
rising only.
Rising is indicat
ed by t he
up
arrow
on the display. To change the test
strategy, press
I
enter"!
,
then use the
®<S>
keys to select the test
strategy.
19