Calibrating a Transmitter
Use the measurement (upper display) and source (lower
display) modes to calibrate a transmitter. The following
example shows how to calibrate a temperature
transmitter.
Connect the calibrator to the instrument under test as
shown in Figure 15. Proceed as follows to calibrate a
transmitter:
1. Press
K for current measurement with loop power.
2. Press
T (lower display). If desired, continue
pressing this key to select the desired thermocouple
type.
3. If necessary, press
M for SOURCE mode.
4. Set your zero and span parameters by pressing X
and W keys. Enter these parameters by pressing and
holding
J and G. For more information on
setting parameters, see Setting 0 % and 100 %
earlier in this manual.
5. Perform test checks at 0-25-50-75-100 % points by
pressing
H or I. Adjust the transmitter as
necessary.

Other manuals for Fluke 724

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 724 and is the answer not in the manual?

Fluke 724 Specifications

General IconGeneral
Temperature Range-10°C to 55°C
Thermocouple TypesJ, K, T, E, R, S, B, L, U, N
Current Measurement0 to 24 mA
Operating Temperature-10°C to 55°C
FunctionProcess calibrator
RTD TypesPt100, Pt500, Pt1000
Voltage Measurement0 to 10 V
Loop Power24 V DC
DisplayDual Display
Power Supply4 AA batteries
Resolution0.1 °C
Storage Temperature-20 °C to 60 °C
Battery Life25 hours
Relative Humidity0 % to 90 %
Safety ComplianceCSA C22.2 No. 1010.1:1992